Steam and water temperature sensing
These assemblies use drilled protective tubes welded and sealed to the equipment, with separate spring-loaded sheathed sensing inserts. The 14 construction serves medium-temperature installations; the 13 construction serves higher-temperature and higher-flow installations. The 01 and 62K constructions use a cylindrical thermal sleeve and a protective tube with three flat faces, allowing hot medium to enter the clearance around the tube.
High-temperature platinum inserts with 650°C capability are available. The operating limits of the assembled configurations are listed separately below.
14, 13 and 01 model operating limits
| Models | Sensor | Elements | Temperature range | Accuracy class | Nominal pressure | Flow velocity | Protective tube material |
| WRNR-14 / WRNR2-14 | K | Single / dual | ≤340°C | I or II | ≤29.4MPa | ≤30m/s | 06Cr18Ni11Ti |
| WRER-14 / WRER2-14 | E |
| WZPR-14 / WZPR2-14 | Pt100 | B |
| WRNR-13 / WRNR2-13 | K | 0-600°C | I or II | ≤100m/s |
| WRER-13 / WRER2-13 | E |
| WZPR-13 / WZPR2-13 | Pt100 | B |
| WRNR-01 / WRNR2-01 | K | I or II |
| WRER-01 / WRER2-01 | E |
| WZPR-01 / WZPR2-01 | Pt100 | B |
62K sensing-tip and operating selections
| Model | Sensor | Measuring junction | Range | Accuracy | Nominal pressure | Flow velocity | Tube material | Response time |
| WRN-624K | K | Insulated | 0-600°C | I or II | ≤30MPa | ≤80m/s | 06Cr18Ni11Ti | ≤60s |
| WRN-625K | Grounded | ≤20s |
| WRE-624K | E | Insulated | ≤60s |
| WRE-625K | Grounded | ≤20s |
| WZP-625K | Pt100 | Not specified | B | Not specified |
14 and 13 protective tube dimensions
| Construction | Head section | Conical section diameters | Sleeve diameter | Insertion selection | Standard exposed length | Head options |
| 14 | 42mm diameter x 50mm length | 32mm / 16mm | Not separately listed | 50mm, 100mm, 150mm, 200mm, 250mm or 300mm | 130mm | Water-resistant aluminum or small polished head |
| 13 | Not separately listed | 38mm / 22mm | 32mm | Choose a total/insertion length pair below | 180mm | Water-resistant aluminum head |
Exposed length can be specified to suit the insulation thickness. Other connection head models are available.
13-type total and insertion length pairs
| Protective tube total length | Insertion length |
| 150mm | 50mm |
| 200mm |
| 100mm |
| 250mm |
| 150mm |
| 300mm |
01 and 62K thermal sleeve dimensions
| Construction | Head section | Cylindrical sleeve | Three-flat section | Insertion | Standard exposed length | Head |
| 01 | 62mm diameter x 50mm length | 28mm diameter x 130mm length | 40mm / 16mm diameters x 120mm length | 120mm | 180mm | Polished head |
| 62K | 24mm diameter x 130mm length | 40mm / 10mm diameters x 120mm length | 150mm |
Alternative exposed lengths are 250mm, 300mm, 350mm, 400mm, 450mm or 500mm, or a specified length. Across these constructions, the single-element electrical entry is M20x1.5 for a 10mm cable; the dual-element entry is M24x1.5 for a 16mm cable.