ELE-802 Submersible Hydrostatic Level Transmitters

Description

Continuous level measurement by immersion

ELE-802 transmitters measure liquid level from hydrostatic pressure. The stainless steel probes are immersed directly in the liquid and provide an electrical signal for a monitoring or control system. Applications include storage and fire-water tanks, wells, reservoirs, municipal water systems, irrigation and compatible wastewater processes.

ModelConfigurationSelection focus
ELE-802General-purpose submersible probeAnalog current, analog voltage or RS485 output; optional companion display
ELE-802TRS485 submersible probeRemote digital monitoring and a shielded vented cable
ELE-802EAnti-clogging probeDense 1mm filter holes for liquids containing particles or sediment
ELE-802LLightning and surge-protected probeOutdoor water monitoring with integrated electrical interference protection

ELE-802 specifications

ParameterELE-802
Measuring range0.4-60m; customizable
Measurement type and mediumGauge pressure; noncorrosive liquids
Housing / diaphragm304 stainless steel / 316L stainless steel
Accuracy±0.5%FS standard; ±0.2%FS or ±0.1%FS optional; ±0.25%FS configuration also available
Overload capacity2× rated range
Outputs and wiring4-20mA, two-wire; RS485, four-wire; 0-5V or 0-10V, three-wire
Supply24VDC nominal; 10-30VDC
ProtectionIP68; three-layer waterproof structure
Working temperature-10°C to 70°C
Humidity0-95%
CableDouble copper shielding
Process connectionsThreaded connection; customized threaded or flange arrangements
CalibrationFactory calibration, aging test and recalibration
Verification referenceJJG 882-2019, Pressure Transmitters

The 3mm pressure guide holes use a 45° angled arrangement to reduce direct contact between sharp objects and the sensing element. A wafer pressure sensor and layered protective construction form the measuring assembly. The bar-graph unit shown with the probe is a separate companion indicator.

Copper cable shielding
Double copper shielding on the ELE-802 signal cable.
Pressure sensing element
Pressure sensing element used for hydrostatic level conversion.
Angled pressure entry holes
Angled pressure entry holes on the probe end.
Layered probe construction
Layered construction protects the probe's internal measuring assembly.

ELE-802T RS485 specifications

ParameterELE-802T
Measurement and installationHydrostatic liquid level; direct submersible installation
OutputRS485
Accuracy±0.5%FS standard; ±0.2%FS or ±0.1%FS optional
ProtectionIP68
Probe housing304 stainless steel standard; 316 stainless steel optional
Diaphragm316L stainless steel
CableVented polyethylene cable with internal pure-copper shielding
Pressure inletThree 3mm holes spaced at 120° intervals
CalibrationHigh-precision equipment calibration and pre-shipment aging test
CustomizationMeasuring range, cable length, materials, accuracy and output configuration
Shielded vented cable construction
The ELE-802T uses a vented polyethylene cable with copper shielding.
Waterproof probe cable connection
Waterproof cable connection for the submersible probe.
High-sensitivity diaphragm assembly
Diaphragm and sensing assembly of the ELE-802T configuration.

ELE-802E anti-clogging specifications

ELE-802E uses a dense pattern of 1mm filter holes to reduce the entry of suspended particles and sediment while transmitting liquid pressure to the sensor. It is intended for compatible sewage, well-water and industrial wastewater applications where clogging is a concern.

Dense filter holes on anti-clogging probe
ELE-802E filter face with densely spaced 1mm holes.
ParameterELE-802E
Measurement principleSubmersible hydrostatic pressure
Measuring range0.2-200mH₂O; customizable to 300m
Output4-20mA
Accuracy±0.5%FS standard; ±0.2%FS or ±0.1%FS optional
Supply24VDC nominal; 10-30VDC
ProtectionIP68
Medium temperature2°C to 50°C
Overload capacity150%FS
Long-term stability±0.1%FS/year
Temperature drift±0.01%FS/°C
Measuring mediumLiquids noncorrosive to the wetted materials
Probe housing304 stainless steel standard; 316 stainless steel optional
Diaphragm316L stainless steel
CableShielded vented cable with braided copper shielding
Electrical connectionEuropean-style terminal block
Signal wiring configurationsTwo-wire current; three-wire voltage; optional four-wire RS485
Integrated circuit board within the probe
Integrated electronics within the ELE-802E probe assembly.
Cable seal and connector detail
Sealed cable connection used with the vented shielded cable.

ELE-802L outdoor and surge-protected specifications

ELE-802L combines a diffused-silicon piezoresistive sensor, a 316L isolation diaphragm and silicone-oil filling. Its integrated protection circuit addresses electrostatic discharge, surge impact and electrical fast transient interference in outdoor monitoring installations. The housing and cable support extended immersion in compatible liquids.

Lightning and surge protection circuit
Integrated protection circuitry in the ELE-802L configuration.
ParameterELE-802L
Measurement typeGauge pressure
Measuring range0.1-100m; customizable
InstallationSubmersible
Measuring mediumLiquids noncorrosive to the wetted materials
Accuracy±0.5%FS standard; ±0.2%FS or ±0.1%FS optional
Output4-20mA standard; RS485 Modbus RTU optional
Supply24VDC nominal; 10-30VDC
Overload capacity150%FS
ProtectionIP68
Housing / diaphragm304 stainless steel / 316L stainless steel
CableDouble-shielded copper cable
Pressure port3mm inlet holes with a 45° angled arrangement
Stability preparationFactory calibration and aging test

Selecting a probe configuration

Match the liquid to the wetted materials, then choose the measuring range, output and cable length. Keep the model-specific overload and temperature ratings with the selected variant. ELE-802E addresses particle-related clogging; ELE-802L adds electrical surge protection; ELE-802T provides the stated RS485 configuration. Choose a compatible external indicator when local numerical or bar-graph display is required.

ELE-802F corrosion-resistant probe

ELE-802F provides a separate corrosion-resistant submersible configuration for liquids compatible with PTFE and ceramic. Its 30mm probe has no threaded process connection and uses a remote cable termination.

ParameterELE-802F
Range selections2–100mH₂O
Pressure referenceGauge
Compatible mediaLiquids compatible with PTFE and ceramic
Probe diameter30mm
Process connectionNone; submersible probe
Output choicesTwo-wire 4–20mA; three-wire 0–5V or 0–10VDC; four-wire RS485
Electrical connectionEuropean terminal, 0.5mm²
4–20mA supply10–30VDC
0–10V supply12–30VDC
Accuracy±0.5%
Overload1.5× full scale standard; special versions available
Temperature0–45°C
ProtectionIP68
Local LCDNone