ELE-801 transmitters use a diffused silicon sensing element and a 304 stainless steel housing for pressure monitoring in water treatment, industrial pipelines, hydraulic and pneumatic equipment, pump stations and HVAC systems. Select wetted materials compatible with the liquid or gas being measured.
The standard connector version, digital display versions, ELE-801Y cable version and ELE-801K circular connector version share the compact threaded construction. Three-point checks at zero, midpoint and full scale, followed by aging tests, support calibration consistency. The pressure-port design reduces pressure shocks; versions rated above 15MPa use an additional damping structure. Laser-marked housing information provides a durable instrument identification.
| Parameter | ELE-801 standard configuration |
|---|---|
| Selectable pressure range | -0.1MPa to 60MPa |
| Accuracy | 0.5%FS |
| Supply | 10–30VDC for the standard current-output configuration |
| Output | 4–20mA; optional 0–5V or 0–10V |
| Process connection | M20×1.5; customization available |
| Operating temperature | 0°C to 50°C |
| Housing | 304 stainless steel |
| Protection | IP65 |
Additional measuring spans and output configurations are available by model. Select a complete measuring range and output; the extended range options below are separate from the standard -0.1MPa to 60MPa configuration.
| Parameter | ELE-801 | ELE-801S | ELE-801Y | ELE-801K | ELE-801ST |
|---|---|---|---|---|---|
| Electrical connection | Hirschmann | Waterproof cable connection | Aviation-style circular connector | Hirschmann | |
| LCD | None | 4 digits | None | 4 digits as standard | |
| Extended range options | ±200–900Pa; -100–900kPa; 1–100MPa | ±200–800Pa; -100–900kPa; 1–100MPa | |||
| Accuracy options | ±0.2% or ±0.5% | ±0.5% | ±0.2% or ±0.5% | ||
| Output configuration | Two-wire 4–20mA; three-wire 0–5V or 0–10VDC; four-wire RS485 | Four-wire RS485 as standard; current or voltage output can be customized for the display head | |||
| Supply for 4–20mA | 10–30VDC | ||||
| Supply for 0–10V | 12–30VDC | 15–30VDC | 12–30VDC | ||
| Process thread options | Male G1/8, G1/4, G3/8, G1/2, G3/4, NPT1/4, NPT3/8, NPT1/2, M10×1, M12×1, M14×1.5 or M20×1.5; customization available | ||||
| Measured media | Liquids and gases compatible with 304 stainless steel | ||||
| Overload | 1.5× full scale as standard; special versions can be customized | ||||
| Temperature | 0°C to 50°C | ||||
| Pressure reference | Gauge or absolute, selected for the application | ||||
| Protection | IP65 | ||||
Current, voltage and RS485 are configuration choices. The ELE-801ST supply entries describe its current and voltage variants; use the supplied nameplate for the RS485 version. Modbus RTU is available with the specified four-wire RS485 configuration.
| Parameter | Configuration options |
|---|---|
| Pressure references | Gauge, absolute, vacuum and compound |
| Gauge-pressure envelope | -0.1 to 100MPa |
| Absolute-pressure envelope | 0 to 100MPa |
| Accuracy | ±0.5%FS standard; ±0.25%FS or ±0.2%FS optional |
| Long-term stability | <0.2%FS per year |
| Wetted isolation diaphragm | 316L stainless steel |
| Housing | 304 stainless steel |
| Local display | LCD on ELE-801S |
| Pressure port | 3mm guide hole; 1.5×FS transient-pressure capability for the specified configuration |
Specify the required pressure reference, calibrated span and accuracy when ordering. The silicon sensor uses an oil-filled isolation assembly, with automatic temperature compensation and signal conditioning.





These reference dimensions describe the illustrated standard bodies. All values are in mm; process thread and connector choices can affect the supplied construction.
| Model | Body diameter | Main body length | Connector height | Display section | Hex size | Shoulder height | Thread length |
|---|---|---|---|---|---|---|---|
| ELE-801 / EM-J10 | 26.5 | 57 | 39.5 | None | 27 | 13 | 18 |
| ELE-801S / EM-J10S | 56 | ||||||
| ELE-801Y | 54 | 17 | None | ||||
| ELE-801K | 37 |
Fit an isolation valve between the transmitter and the process. Depressurize and disconnect power before installation or removal, and apply wrench torque to the hexagonal fitting. Increase pressure gradually, fit a snubber for rapid pressure changes and use a suitable cooling arrangement when the medium exceeds the configured temperature rating. Mount ranges below 30kPa vertically.
For the Hirschmann two-wire current version, pin 1 is loop positive and pin 2 is loop negative. For the three-wire voltage version, pin 1 is supply positive, pin 2 is common negative and pin 3 is signal positive. Use the configuration-specific diagram in the English manual for RS485 and other connector versions. Route shielded signal cable separately from power wiring and keep moisture out of the cable entry.
The attached English catalogue includes the full transmitter range. The English user manual contains selection guidance, dimensions, connector and cable assignments, current-loop and voltage circuits, RS485 connections, installation precautions and service information.